Contrast enhancement of spectral domain optical coherence tomography using spectrum correction
Guangming Ni1, Linbo Liu2, Xiaojun Yu3
1State Key Laboratory of Electronic Thin Films and Integrated Devices, School of Optoelectronic Information, University of Electronic Science and Technology of China, No. 4, Section 2, North Jianshe Road, Chengdu, 610054, China; School of Electrical and Electronic Engineering, Nanyang Technological University, 639798, Singapore.
Abstract:
We report a spectrum correction method to enhance the image contrast of spectral domain optical coherence tomography (SD-OCT). Our method treats SD-OCT signals as the product of harmonic signals backscattered from a sample comprising a series of discrete reflectors and a window corresponding to the light source spectrum. The method restores the magnitude of the main lobe of the axial point spread function (PSF) by estimating the magnitudes of the backscattered harmonic signals and strengthens OCT signals using these estimated values. Experimental results acquired from fresh rat corneas and fixed human aortic atherosclerosis tissues show that our method provides clearer microstructural information than the conventional methods by improving the contrast to noise ratios (CNRs) by 1.4779 dB and 3.2595 dB, respectively. This improved image quality is obtained without any hardware change, making our method a cost-effective alternative to compete with hardware advances.
More Related Videos
12:54Simultaneous Brightfield, Fluorescence, and Optical Coherence Tomographic Imaging of Contracting Cardiac Trabeculae Ex Vivo
Published on: October 2, 2021
09:56Automated 3D Optical Coherence Tomography to Elucidate Biofilm Morphogenesis Over Large Spatial Scales
Published on: August 21, 2019
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Aliasing
If the sampling frequency is below the Nyquist rate, these replicas overlap, preventing the original...
Attenuated Total Reflectance (ATR) Infrared Spectroscopy: Overview
The ATR process begins by directing a beam...
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...