Knowing About Device Algorithms to Understand the Rhythm and Role of Managed Ventricular Pacing
Antoine Kossaify1, Hiba Hamdar2
1Electrophysiology Unit, Cardiology Division, USEK-University Hospital Notre Dame des Secours, Byblos, Lebanon.
Abstract:
A 66-year-old patient with a DR pacemaker for intermittent atrioventricular block presented with an electrocardiogram (ECG) showing some P waves non followed by QRS complexes, raising suspicion of device dysfunction. The device was equipped with a special algorithm (Managed Ventricular Pacing; Medtronic), and the observed ECG tracing was a normal consequence of the function of such algorithms. Being aware of the function of specific algorithms is essential to adequately analyze rhythms.
More Related Videos
11:00Assessment of Sarcoplasmic Reticulum Calcium Reserve and Intracellular Diastolic Calcium Removal in Isolated Ventricular Cardiomyocytes
Published on: September 18, 2017
08:19Electrophysiological Assessment of Murine Atria with High-Resolution Optical Mapping
Published on: February 22, 2018
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Dysrhythmias VI: Management of Dysrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Electrophysiology of Normal Cardiac Rhythm
Dysrhythmias I: Introduction
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....