Identifying Coronary Artery Calcification on Non-gated Computed Tomography Scans
Published on: August 28, 2018
Estimation and comparison of the radiation effective dose during coronary computed tomography angiography
Daryoush Khoramian1, Soroush Sistani2
1Independent Researcher, Tehran, Iran.
Goal:
To estimate and compare the radiation dose associated with coronary computed tomography angiography (CCTA) examinations on two multi-detector CT scanners (MDCT), 64-MDCT and 128-MDCT, in daily practice.
Methods:
Scan parameters of 90 patients undergoing retrospective electrocardiographic gating spiral CCTA exam were recorded during a period on a single-source 64-MDCT and a dual-source 128-MDCT, and average scan parameters were derived that were used for dosimetry. The computed tomography dose index (CTDI) with a pencil ionisation chamber and polymethyl methacrylate body phantom with diameter of 32 cm was measured on both scanners. The dose-length product (DLP) was calculated and the DLP to effective dose conversion factor (for chest scan at 120 kV of 0.014 mSv mGy-1 cm-1) was used to estimate effective dose (ED).
Results:
Patients' heart rate, scan length, pitch factor, CTDIv, DLP and ED for 128-MDCT were 64 (5) (beats min-1), 161 (10) (mm), 0.26, 47 (12) (mGy), 769 (212) (mGy cm) and 10.3 (3.1) (mSv), respectively [mean (one standard deviation)]. Patients' heart rate, scan length, pitch factor, CTDIv, DLP and ED for 64-MDCT were 60 (7) (beats min-1), 172 (14) (mm), 0.2, 60 (6) (mGy), 1068 (98) (mGy cm) and 14.9 (1.4) (mSv), respectively.
Conclusion:
Our results indicated that the CTDIv, DLP and the effective dose with 128-MDCT is significantly lower than with 64-MDCT (p < 0.05). As differences between the exposure parameter mAs on two CT scanners was not significant (p > 0.05) and the kV was constant for both scanners (120 kV), the differences resulted from a shorter scan length on the 128-MDCT and use of a higher pitch factor (0.26 and 0.2 in the 128-MDCT and 64-MDCT, respectively). Comparison with other published studies confirms the findings and indicates methods for reducing patient dose.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
11:38Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies III: Computed Tomography