Related Experiment Video
Updated: Feb 22, 2026

Calorespirometry: A Powerful, Noninvasive Approach to Investigate Cellular Energy Metabolism
Published on: May 31, 2018
Continuous measurement of oxygen consumption using the reversed fick method
Michihiko Fukui1, Maho Imoto2, Nobuaki Shime2
1Intensive Care Unit, Otsu Municipal Hospital, 2-9-9 Motomiya, 520, Otsu, Shiga, Japan.
Abstract:
We developed a continuous oxygen consumption (Vo2) measurement system employed the reversed Fick method, in which Vo2 in computed from continuously measured sured arterial and mixed venous oxygen saturation assed by pulse oximetry and mixed venous oximetry, respectively, and cardiac output by the heat deprivation technique. This system was compared with the conventional intermittent reversed fick method in 7 patients during surgery and with indirect calorimetry in 4 intensive care unit (ICU) patients. The Vo2 measured by the continuous reversed Fick method showed a high correlation with those simultaneously measured by the intermittent Fick method (r=0.97,P<0.01) and by indirect calorimetry (r=0.74,P<0.01). The 95% confidence limits (bias±2 SD) of the continuous reversed Fick method were -0.6±45 ml·min-1 with the intermittent Fick method and -31±56 ml·min-1 with indirect calorimetry. The continuous Fick method is in satisfactory agreement with the conventional methods for the measured of Vo2 and potentially allows for convenient assessment of Vo2 in critically ill patients.
More Related Videos
07:09Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
07:12Author Spotlight: Influence of Temperature on Drosophila melanogaster and Desert-Adapted Beetles
Published on: July 7, 2023
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...