Robust and practical non-invasive estimation of local arterial wave speed and mean blood velocity waveforms
Remi Kowalski1,2,3, Richard Beare4,5, Marie Willemet6
1Heart Research, Clinical Sciences, Murdoch Childrens Research Institute, Parkville, Victoria, Australia.
Objective:
Local arterial wave speed, a surrogate of vessel stiffness, can be estimated via the pressure-velocity (PU) and diameter-velocity (ln(D)U) loop methods. These assume negligible early-systolic reflected waves (RWes) and require measurement of cross-sectionally averaged velocity (U mean), which is related to volumetric blood flow. However, RWes may not always be negligible and Doppler ultrasound typically provides maximum velocity waveforms or estimates of mean velocity subject to various errors (U raw). This study investigates how these issues affect wave speed estimation and explores more robust methods for obtaining local wave speed and U mean.
Approach:
Using aortic phase-contrast MRI (PCMRI, n = 34) and a simulated virtual cohort (n = 3325), we assessed errors in calculated wave speed caused by RWes and use of U raw rather than true U mean. By combining PU raw and ln(D)U raw loop wave speed values, (i) a corrected wave speed (ln(D)P), insensitive to RWes and velocity errors, was derived; and (ii) a novel method for estimating U mean from U raw was proposed (where U raw can be any scaled version of U mean).
Main Results:
Proof-of-principle was established via PCMRI data and in the ascending aorta, carotid, brachial and femoral arteries of the virtual cohort, with acceptably low wave speed and U mean errors obtained even when local pressure was estimated from diameter and mean/diastolic brachial pressures.
Significance:
Given a locally measured diameter waveform and brachial cuff pressures, (i) the velocity- and RWes-independent ln(D)P method can be applied non-invasively and is likely more robust than ln(D)U and PU loop methods; and (ii) U mean can be estimated from routinely-acquired U raw.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure