Thoracic endovascular aortic repair migration and aortic elongation differentiated using dual reference point
Hillary B Alberta1, Toshio Takayama1, Annalise Panthofer1
1Division of Vascular Surgery, School of Medicine and Public Health, University of Wisconsin, Madison, Wisc.
Journal of Vascular Surgery
|September 26, 2017
Summary
Dual reference points accurately differentiate aortic elongation from true stent graft migration in thoracic endovascular aortic repair. This method improves diagnostic accuracy by accounting for natural aortic length changes over time.
Area of Science:
- Vascular Surgery
- Medical Imaging
- Biomedical Engineering
Background:
- Thoracic endovascular aortic repair (TEVAR) is a common procedure for aortic pathologies.
- Stent graft migration is a potential complication, but distinguishing it from aortic elongation can be challenging.
- Conventional migration definitions may lead to false positives due to natural aortic lengthening.
Purpose of the Study:
- To evaluate the efficacy of using dual reference points for differentiating stent graft migration from aortic elongation after TEVAR.
- To compare the accuracy of dual reference point measurements against conventional single reference point methods.
Main Methods:
- Retrospective analysis of 23 patients who underwent TEVAR with the GORE TAG thoracic endoprosthesis.
- Measurement of aortic centerline lengths using innominate artery (IA) to distal ring, IA to celiac artery (CA), and distal ring to CA intervals.
- Comparison of baseline measurements with interval measurements meeting standard migration criteria.
Main Results:
- Twenty of 23 patients (87%) exhibited aortic elongation, not true migration, when assessed with dual reference points.
- Significant differences in IA-CA and distal ring-CA interval length changes were observed between elongation and migration groups (P < .05).
- The IA to distal ring distance remained similar between elongation and migration, highlighting its limitation as a single marker.
Conclusions:
- Dual reference point imaging is crucial for accurately distinguishing aortic elongation from true stent graft migration.
- Accounting for dynamic aortic length changes improves diagnostic precision in TEVAR follow-up.
- This approach helps avoid misdiagnosis of migration, leading to more appropriate patient management.