Related Experiment Video
Updated: Feb 21, 2026

Three-dimensional Optical-resolution Photoacoustic Microscopy
Published on: May 3, 2011
Correcting the limited view in optical-resolution photoacoustic microscopy
Wei Liu1, Yuan Zhou1,2, Mengran Wang3
1Department of Biomedical Engineering, Duke University, Durham, North Carolina.
Abstract:
Optical-resolution photoacoustic microscopy (OR-PAM) has proven useful for anatomical and functional imaging with high spatial resolutions. However, the coherent signal generation and the desired reflection-mode detection in OR-PAM can result in a limited detectability of features aligned with the acoustic axis (ie, vertical structures). Here, we investigated the limited-view phenomenon in OR-PAM by simulating the generation and propagation of the acoustic pressure waves and determined the key optical parameters affecting the visibility of vertical structures. Proof-of-concept numerical experiments were performed with different illumination angles, optical foci and numerical apertures (NA) of the objective lens. The results collectively show that an NA of 0.3 can readily improve the visibility of vertical structures in a typical reflection-mode OR-PAM system. This conclusion was confirmed by numerical simulations on the cortical blood vessels in a mouse brain and by experiments in a suture-cross phantom and in a mouse brain in vivo.