A novel non-invasive blood pressure waveform measuring system compared to Millar applanation tonometry
Sándor Földi1, Tamás Horváth2, Flóra Zieger3
1Faculty of Information Technology and Bionics, Pázmány Péter Catholic University, Budapest, Hungary. foldi.sandor@itk.ppke.hu.
Abstract:
The accurate, non-invasive, measuring of the continuous arterial blood pressure waveform faces some difficulties and an innovative blood pressure measurement technology is urgently needed. However, the arterial blood pressure waveform plays an essential role in health care by providing diagnostic information and base for calculating several heart function parameters. The aim of this study is to introduce a novel non-invasive measuring system that can measure the arterial blood pressure waveform with high accuracy in comparison to an applanation tonometry system. The applied measuring device utilizes a new measurement strategy enabled by the OptoForce 3D force sensor, which is attached to the wrist at the radial artery. To validate the accuracy, 30 simultaneous measurements were taken with a Millar tonometer. For the simultaneously recorded non-invasive signals, the similarity was high (the average correlation was [Formula: see text]). The differences in the systolic and the diastolic blood pressure measured by the two systems are small. The average differences ([Formula: see text]) for simultaneously recorded systolic, diastolic, mean arterial and incisura pressures were: [Formula: see text], [Formula: see text], [Formula: see text] and [Formula: see text], respectively. These results satisfy the AAMI criteria. Based on our results, this new system requires further development and validation against invasive arterial blood pressure monitoring in order to prove its usefulness in patient monitoring, emergency care, and pulse diagnosis.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement