COMPARISON OF PERSONAL DOSE EQUIVALENT Hp(10) IN 137CS RADIATION BETWEEN THE PRIMARY STANDARDS LABORATORIES OF JAPAN
D J Butler1, T Kurasawa2, M Litwin1
1Australian Radiation Protection and Nuclear Safety Agency (ARPANSA), Yallambie, Victoria 3085, Australia.
Japan and Australia compared personal dose equivalent Hp(10) using BeO OSL dosemeters for 137Cs radiation. Results confirmed high precision, with a ratio of 1.006, validating passive dosimetry for standards comparisons.
Area of Science:
- Radiation Dosimetry
- Metrology
- Applied Physics
Background:
- Accurate measurement of personal dose equivalent (Hp(10)) is crucial for radiation protection.
- International comparisons of dosimetry standards ensure consistency and reliability.
- Passive Optically Stimulated Luminescence (OSL) dosemeters offer a viable method for such comparisons.
Purpose of the Study:
- To compare personal dose equivalent (Hp(10)) measurements for 137Cs radiation between Japan and Australia's primary standards laboratories.
- To evaluate the precision achievable with commercially available BeO OSL dosemeters.
- To verify the personal dose equivalent delivery methodologies employed by each standards laboratory.
Main Methods:
- Utilized 120 passive BeO OSL dosemeters (Dosimetrics GmbH).
- Delivered a 5 mSv dose of 137Cs radiation to 40 dosemeters in each participating country.
- Included 40 dosemeters as controls in each country for baseline assessment.
Main Results:
- The ratio of Hp(10) measurements between Japan and Australia was determined to be 1.006.
- A combined standard uncertainty of 3.2% was achieved for the comparison.
- Statistical uncertainty was as low as 0.32%, demonstrating high precision.
Conclusions:
- Passive BeO OSL dosemeters are suitable for high-precision dosimetry comparisons.
- The study confirmed the accuracy and consistency of personal dose equivalent delivery methods in both laboratories.
- The findings support the use of OSL dosimetry for inter-laboratory metrological comparisons.
More Related Videos
09:03Radiotracer Administration for High Temporal Resolution Positron Emission Tomography of the Human Brain: Application to FDG-fPET
Published on: October 22, 2019
07:28Lipidico Injection Protocol for Serial Crystallography Measurements at the Australian Synchrotron
Published on: September 23, 2020
Related Concept Videos
Biological Effects of Radiation
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Atomic Emission Spectroscopy: Overview
Atomic Emission Spectroscopy: Lab