Quantification accuracy of neuro-oncology PET data as a function of emission scan duration in PET/MR compared to
Stefan Wampl1, Ivo Rausch2, Tatjana Traub-Weidinger3
1QIMP group, Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Waehringer Guertel 18-20, 1090 Vienna, Austria; Sensors and Ultrasonics group, Institute of Applied Physics, Vienna University of Technology, Wiedner Hauptstraße 8-10/E134, 1040 Vienna, Austria.
Objectives:
To evaluate and compare the effect of reduced acquisition time, as a surrogate of injected activity, on the PET quantification accuracy in PET/CT and PET/MR imaging.
Methods:
Twenty min 18F-FDG phantom measurements and 10min 18F-FET brain scans were acquired in a Biograph-True-Point-True-View PET/CT (n=8) and a Biograph mMR PET/MR (n=16). Listmode data were repeatedly split into frames of 1min to 10min length and reconstructed using two different reconstruction settings of a 3D-OSEM algorithm: with post-filtering ("OSEM"), and without post-filtering but with resolution recovery ("PSF"). Recovery coefficients (RCmax, RCA50) and standard uptake values (SUVmax, SUVA50) were evaluated.
Results:
RCmax (phantom) and SUVmax (patients) increased significantly when reducing the frame duration. Significantly lower deviations were observed for RCA50 and SUVA50, respectively, making them more appropriate to compare PET studies at different number of counts. No statistical significant differences were observed when using post-filtering and reducing the frame time to 4min (RCA50, reference 20min, phantom) and to 3min (SUVA50, reference 10min, patients).
Conclusions:
For hybrid aminoacid brain imaging, frame duration (or injected activity) can potentially be reduced to 30% of the standard used in clinical routine without significant changes on the quantification accuracy of the PET images if adequate reconstruction settings and quantitative measures are used. Frame times below 4min in the NEMA phantom are not advisable to obtain quantitative and reproducible measures.