Characterization of Artifacts Associated With Multicolor Confocal Scanning Laser Ophthalmoscopy
Background And Objective:
To characterize the appearance of three types of artifacts observed on multicolor confocal scanning laser ophthalmoscopy (cSLO).
Patients And Methods:
Retrospective review of 159 eyes of 96 consecutive patients from the Duke Eye Center who underwent multicolor cSLO with spectral-domain optical coherence tomography (SD-OCT). Infrared (IR), green, blue, and multicolor reflectance images were evaluated for artifacts with corresponding SD-OCT scans available for reference.
Results:
Multicolor cSLO artifacts were detected in 23.3% (37 of 159) of eyes and comprised three main patterns: spot, wisp, and net. Only three instances of these artifacts were detected on IR reflectance versus 34, 37, and 35 instances on green, blue, and multicolor reflectance, respectively. Artifacts were observed in 0% of eyes with clear lenses, 27.7% of eyes with cataracts, and in 20.8% of eyes with posterior chamber intraocular lenses.
Conclusion:
Awareness of spot, wisp, and net artifacts when interpreting multicolor cSLO images may facilitate the identification of true retinal pathology. [Ophthalmic Surg Lasers Imaging Retina. 2017;48:810-815.].