Related Experiment Videos
Novel technique for intraoperative sciatic nerve assessment in complex primary total hip arthroplasty: a pilot study
Shai S Shemesh1, Jonathan Robinson1, Samuel Overley1
1Department of Orthopaedic Surgery, Icahn School of Medicine at Mount Sinai, New York - USA.
Summary
A handheld nerve stimulator effectively monitored sciatic nerve (SN) function during complex total hip arthroplasty (THA), preventing permanent nerve injury. This tool aided surgical decisions, allowing immediate adjustments to avoid sciatic nerve injury (SNI).
Area of Science:
- Orthopedic Surgery
- Neurophysiology
- Medical Devices
Background:
- Sciatic nerve injury (SNI) is a significant risk during total hip arthroplasty (THA), especially in complex cases.
- Intraoperative neural monitoring can prevent SNI but is often challenging to implement.
- This study evaluates a handheld nerve stimulator for SN monitoring in complex THA.
Purpose of the Study:
- To assess the efficacy of a handheld nerve stimulator for intraoperative sciatic nerve (SN) monitoring during complex total hip arthroplasty (THA).
- To determine if this monitoring technique can prevent sciatic nerve injury (SNI) in cases requiring significant limb lengthening or nerve manipulation.
Main Methods:
- A series of 11 complex THA cases (Crowe 3-4 hip dysplasia or similar) were analyzed.
- Sciatic nerve (SN) function was monitored using a handheld nerve stimulator, measuring pre- and post-reduction thresholds.
- Nerve stimulation results guided intraoperative decision-making.
Main Results:
- No permanent sciatic nerve injury (SNI) occurred in any of the 11 cases.
- A handheld nerve stimulator identified two instances where surgical plans were altered (femoral shortening, smaller femoral head) to prevent potential nerve damage.
- One patient experienced a transient peroneal nerve palsy that resolved within two weeks.
Conclusions:
- Intraoperative use of a handheld nerve stimulator is effective in guiding surgical decisions during complex THA.
- This technique can potentially prevent sciatic nerve injury (SNI) by allowing real-time assessment and corrective action.
- The nerve stimulator provides valuable feedback to ensure safe limb lengthening and nerve manipulation.