Inkjet-printed barcodes for a rapid and multiplexed paper-based assay compatible with mobile devices
Mingzhu Yang1, Wei Zhang, Wenshu Zheng
1Beijing Engineering Research Center for BioNanotechnology and CAS Key Lab for Biological Effects of Nanomaterials and Nanosafety, CAS Center for Excellence in Nanoscience, National Center for Nanoscience and Technology, Beijing 100190, China. zhangw@nanoctr.cn xingyujiang@nanoctr.cn.
This study introduces a rapid, low-cost barcoded paper-based assay (BPA) for multiplexed detection. The system uses a mobile app for quick analysis, ideal for on-site monitoring of drug residues and infections.
Area of Science:
- Analytical Chemistry
- Biosensors
- Point-of-Care Diagnostics
Background:
- Traditional diagnostic methods can be slow, expensive, and require specialized equipment.
- There is a need for rapid, low-cost, and portable diagnostic tools for on-site testing.
Purpose of the Study:
- To develop a simple, rapid, low-cost, and multiplexed barcoded paper-based assay (BPA) compatible with mobile devices.
- To enhance coding capacity for multiplexed detection using novel barcodes and a dedicated mobile application.
- To demonstrate the BPA system's utility in detecting drug residues and other analytes.
Main Methods:
- Mass-manufacturing of barcoded paper-based analytical devices (BPADs) using inkjet printing and XYZ dispensing.
- Design of a new barcode group with significantly higher coding capacity compared to standard codes.
- Development of a mobile application for efficient reading and interpretation of the new barcodes.
- Validation of the BPA system for drug residue detection, specifically enrofloxacin (ENR).
Main Results:
- Achieved high precision and efficiency in BPAD mass-manufacturing.
- The new barcodes demonstrated 16 times higher coding capacity than Codabar.
- The BPA system enabled complete detection and result readout within 10 minutes.
- The limit of detection for ENR was 8 ng mL-1, meeting monitoring requirements.
Conclusions:
- The developed BPA system is simple, rapid, low-cost, robust, and multiplexed.
- The system's high coding capacity and mobile compatibility make it suitable for various applications, including blood-transmitted infections and drug residue monitoring.
- The BPA system is highly suitable for rapid and on-site detection due to its efficiency and selectivity.