Introduction:
Ferromagnetic port-containing breast tissue expanders are currently labeled MRI-unsafe because of the presumption that magnets should not enter the machine. However, designating these devices as MRI-unsafe can lead to unnecessary procedures or suboptimal imaging choices. This study provides an ex vivo analysis of how breast tissue expanders behave when subjected to strong magnetic fields to determine which variables might affect clinical risk.
Methods:
Three different brands of tissue expanders were evaluated in three MRI environments. Translational force was determined using the deflection angle method. Torque on empty, saline-filled, and air-filled expanders was evaluated on a 0-4 scale. Magnetic field was measured using a gaussmeter. The weight required to prevent displacement of the expanders was determined for both air- and saline-filled expanders. Temperature over time was measured using an alcohol thermometer.
Results:
Magnetic field strength, deflection angle, and torque were the greatest in 3T MRI environments and varied by device manufacturer (Sientra > Mentor > Allergan). Saline-filled expanders required 240 mL and air-filled required 360 mL volume to make the torque undetectable, and the effect of torque could be mitigated with prone positioning. A weight of 120 g was required to prevent displacement of a saline-filled tissue expander and 870 g for an empty expander. There were no appreciable changes in temperature.
Conclusions:
Previously described risks may be reduced by using a 1.5T MRI, device selection, filling expanders with saline, and prone positioning. MRI can be considered in patients with breast tissue expanders when appropriate peri-procedural choices have been made so that the benefits of undergoing MRI outweigh the risks.
More Related Videos
12:18Magnetic Resonance Elastography Methodology for the Evaluation of Tissue Engineered Construct Growth
Published on: February 9, 2012
13:35Endoscopic Bilateral Nipple-sparing Mastectomy via a Single Axillary Incision with Immediate Pre-pectoral Implant-based Breast Reconstruction
Published on: May 17, 2024
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies IV: Magnetic Resonance Imaging
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...