Related Experiment Video
Updated: Feb 20, 2026

Ion-Exchange Membranes for the Fabrication of Reverse Electrodialysis Device
Published on: July 20, 2021
Verification of physicochemical structures of dialysis membrane using reversal dialysis technique
Akihiro C Yamashita1, Takahisa Ono1, Narumi Tomisawa1
1Department of Chemical Science and Technology, Faculty of Bioscience and Applied Chemistry, Hosei University, Tokyo, Japan.
Objective:
Solute removal performance of the dialyzer is closely related to physicochemical structures of the membrane. The objective of this study is to devise a new in vitro evaluation technique to directly correlate the physicochemical structures of the membrane to the dialysis performance.
Materials And Methods:
Commercial 11 dialyzers with cellulose triacetate (CTA), ethylene vinylalcohol co-polymer (EVAL), asymmetry CTA (termed ATA), 4 polysulfone (PSf), polyether sulfone (PES), and 3 polyester polymer alloy (PEPA) membranes were employed for investigation. Normal dialysis experiments were performed with aqueous creatinine (MW 113), bromophenol blue (MW 670), vitamin B12 (MW 1355), chymotrypsin (MW 25,000) test solutions. Reversal dialysis experiments were devised introducing the test solution outside the hollow fiber (HF) and the dialysate inside the HF, respectively. Clearances were measured under QB = 200 mL/min and QD = 500 mL/min at 310 K. The ratio of clearance in the reversal dialysis to that in the normal dialysis termed the yielding factor of asymmetry, fYA , was defined for evaluation.
Results And Discussion:
According to the classic mass transfer theory, clearance in the normal dialysis and that in the reversal dialysis are expected to be identical; however, the average fYA was 0.96 for creatinine and was always slightly lower than unity in 10 dialyzers out of 11. This may be caused by un-uniform distribution of the test solution outside the HF, which was more likely to occur than when it flowed inside the HF. Unlike creatinine, the average fYA for vitamin B12 was 1.0 and that for chymotrypsin was 1.06 to 1.45 in membranes with heterogeneous structures (ATA, PSf, PEPA, and PES), but was almost unity (1.02) in homogeneous membrane (CTA). This must be based on the fact that the membrane permeability in the reversal dialysis is much larger than that in the normal dialysis due to the physicochemical structures of the heterogeneous membrane that has a wedge-like pore size distribution in the radial direction.
Conclusions:
A wedge-like pore size distribution in the radial direction in heterogeneous membrane was semiquantitatively evaluated by introducing a reversal dialysis technique with a new index.
Related Concept Videos
Dialysis
Dialysis
Acute kidney injury develops suddenly and can be caused by pre-renal causes (e.g., hypovolemia, shock), intrinsic renal causes (e.g., acute tubular necrosis), or post-renal causes (e.g., urinary obstruction). In contrast, chronic renal failure progresses gradually over time and is often...
Peritoneal Dialysis I: Introduction and Procedure
Hemodialysis II: Procedure and Complications
Detergent Purification of Membrane Proteins
Hemodialysis I: Introduction