Visual assessment of hemolysis affects patient safety
Ana Helena Luksic1, Nora Nikolac Gabaj2, Marijana Miler2
1Department of Clinical Chemistry, Sestre Milosrdnice University Hospital Center, Vinogradska cesta 29, Zagreb, Croatia, Phone: +385996913474.
Background:
Manual handling of hemolyzed samples is not standardized and is vulnerable to errors. This study aimed to evaluate laboratory errors due to manual handling of hemolyzed samples and to assess the risk they might have for patient safety.
Methods:
Data were retrospectively obtained from a laboratory information system for 25 emergency tests from hemolyzed samples. Hemolysis (concentration of free hemoglobin >0.5 g/L) was visually assessed by comparison with a color chart. The reference person reestimated the routinely assessed degree of hemolysis to all samples (n=3185) received in the laboratory in a 1-week period. For each test, the correct and incorrect way of handling results was determined. Risk assessment was performed according to ISO 14971 standard with five categories of risk (S1-S5) and error occurrence (O1-O5).
Results:
In the studied period, the emergency laboratory received 495 hemolyzed samples (15.5%) with a total of 2518 laboratory test requests (15.5%): 102 (20.6%) of the reports from hemolyzed samples had a comment on hemolysis; 31% of the test results were handled incorrectly (20.7% due to the incorrect release of the test result despite hemolysis interference and 10.3% due to unnecessary suppression), accounting for 4.8% of the total test volume. Tests with the highest combination of risk and occurrence rate were troponin T, potassium and total bilirubin.
Conclusions:
Manual handling of hemolyzed samples may lead to risk of errors in reporting results for troponin T, potassium and total bilirubin, which may have an effect on clinical decision. In addition, unnecessary suppression of the sample results unaffected by hemolysis could affect patient outcome.
More Related Videos
08:44An In Vitro Hemodynamic Loop Model to Investigate the Hemocytocompatibility and Host Cell Activation of Vascular Medical Devices
Published on: August 21, 2020
07:24A Precision Medicine Tool for Measurement and Monitoring of Hemoglobin S in Sickle Cell Disease Patients Receiving Transfusion Therapy
Related Concept Videos
Changes in Skin Color: Clinical Perspectives
Albinism
Albinism is a genetic disorder that affects (completely or partially) the coloring of skin, hair, and eyes. The defect is primarily...
Assessment of the Cardiovascular System II: Inspection
Head and Neck
Effect of Hepatic Disease on Pharmacokinetics: Pathophysiologic Assessment and Liver Function Test