TOPPE: A framework for rapid prototyping of MR pulse sequences
Jon-Fredrik Nielsen1, Douglas C Noll1
1Department of Biomedical Engineering, University of Michigan, Ann Arbor, Michigan, USA.
A new file format, TOPPE (Targeted Optimized Pulse-sequence Prototyping Environment), enables rapid prototyping of magnetic resonance (MR) pulse sequences. This framework simplifies MR imaging experiment design, reducing the need for hardware-specific programming.
Area of Science:
- Magnetic Resonance Imaging
- Medical Physics
- Biomedical Engineering
Background:
- Developing novel magnetic resonance (MR) pulse sequences is crucial for advancing imaging capabilities.
- Traditional MR pulse sequence programming requires specialized hardware knowledge and extensive compilation, hindering rapid innovation.
Purpose of the Study:
- To introduce a novel framework, the Targeted Optimized Pulse-sequence Prototyping Environment (TOPPE), for the rapid prototyping of MR pulse sequences.
- To simplify the process of designing and executing MR imaging experiments by abstracting hardware-specific details.
Main Methods:
- A simple file format, TOPPE, was developed to specify all aspects of an MR imaging experiment, including gradient and radiofrequency waveforms and scan loops.
- A GE scanner-compatible binary executable (interpreter) was created to load and execute TOPPE files.
- MATLAB scripts were provided for creating, reading, writing, and previewing TOPPE files before hardware execution.
Main Results:
- The TOPPE framework successfully implemented and executed three distinct MR sequences: k-space mapping, non-Cartesian PRESTO whole-brain dynamic imaging, and myelin mapping.
- A single TOPPE interpreter binary was utilized to execute multiple sequences by simply modifying TOPPE files, demonstrating flexibility.
- The need for sequence-specific compilation was eliminated, with the interpreter requiring compilation only once per scanner software upgrade.
Conclusions:
- TOPPE provides a complete specification for MR imaging experiments using a modular approach with unique modules.
- The TOPPE file format, coupled with the GE interpreter, establishes a flexible platform for accelerating the development and prototyping of new MR pulse sequences.
- This approach significantly reduces the barrier to entry for pulse sequence programming, fostering innovation in MR imaging research.