Related Experiment Video
Updated: Feb 19, 2026

A Standardized Method for Measurement of Elbow Kinesthesia
Published on: October 10, 2020
An X-ray-free method to accurately identify the elbow flexion-extension axis for the placement of a hinged external
Jian Song1, Hui Ding1, Wei Han2
1Department of Biomedical Engineering, School of Medicine, Tsinghua University, Room C249, Beijing, 100084, People's Republic of China.
Purpose:
Identifying the elbow flexion-extension (F-E) movement axis is important for placing a hinged elbow external fixator. An X-ray fluoroscopy-based method is widely used in clinical practice, exposing the patient and surgeons to high doses of radiation. Additionally, the accuracy and repeatability of the fluoroscopy-based method are very low and affected by subjective factors.
Methods:
To solve this problem, an X-ray-free method based on kinematics analysis was proposed to identify the elbow F-E movement axis, and a navigation system was built to guide the placement of the elbow external fixator.
Results:
Our X-ray-free navigation method is more repeatable than the current X-ray fluoroscopy method used clinically. Both our algorithm and the NIST (National Institute of Standards and Technology) algorithm showed high accuracy and repeatability to identify the axis.
Conclusions:
The method proposed in this study is very promising to avoid a large dose of X-ray radiation and increases the repeatability and performance of identifying the elbow F-E movement axis for the placement of the hinged elbow external fixator.