Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
12.2K
Standardization of 67Cu and calibration of the ionization chamber. Impurities and decay scheme problems
Abstract:
This paper describes the standardization of 67Cu by the 4πβ(PC)-γ coincidence method, and the calibration of ionization chamber. The difficulties were. (i) One of the excited levels of 67Zn has a half life of 9.10μs. (ii) A recent publication reconsiders the decay scheme parameters. (iii) The solution had a significant content of impurities. The conclusions were that the result of absolute standardization is less influenced by the decay scheme parameters and impurities than the measurements by ionization chamber.