Related Experiment Video
Updated: Feb 19, 2026

Structured Approach to Colonoscopy Technique Optimization: A Single-Center Experience with Novice Endoscopists
Published on: July 11, 2025
A novel device for improving visualization in an inadequately prepared colon
Seth A Gross1, Lauren B Gerson2, Blair S Lewis3
1Tisch Hospital, NYU Langone Medical Center, New York, New York, USA.
Background And Aims:
Adequate bowel preparation is crucial for a quality colonoscopy. Poor bowel preparation can result in longer procedural times, missed adenomas, earlier repeat procedures, increase costs, and decreased patient satisfaction. Pure-Vu (MOTUS GI, Tirat Carmel, Israel) is a system that facilitates intraprocedural cleaning of a poorly prepared colon during a colonoscopy by irrigating the colon and evacuating the fluid and feces. The study aims were to evaluate Pure-Vu's cleansing capabilities and its usability and safety in a live animal colon.
Methods:
The Pure-Vu system was used by 4 experienced gastroenterologists in 35 Yorkshire cross swine (66% female) that received a reduced bowel preparation to ensure an inadequate bowel preparation at baseline. Before colonoscopy the Pure-Vu was attached to the colonoscope, and the baseline prep was assessed during insertion. The Pure-Vu system was then used to cleanse the colon and the prep then assessed after Pure-Vu use.
Results:
No adverse effects and no failed or prematurely terminated cases were noted. Fourteen percent of the swine colons were adequately prepped at baseline (median Boston Bowel Preparation Scale [BBPS] score = 0 [25th percentile = 0; 75th percentile = 1; interquartile range = 1]) and improved to 100% after use of Pure-Vu (Wilcoxon signed rank test P < .001) (median BBPS score = 3 [25th percentile = 3; 75th percentile = 3; interquartile range = 0]). The physicians found Pure-Vu easy and intuitive to operate.
Conclusions:
The Pure-Vu system effectively cleaned inadequately prepped swine colons and proved to be easy to use. The Pure-Vu device, applied in clinical practice, may result in improved overall prep quality and a reduction in repeat examinations because of inadequately prepped colons.
Insights
The Pure-Vu system effectively cleans poorly prepared colons during colonoscopy in animal models. This innovative device improves bowel prep quality, potentially reducing repeat procedures and enhancing patient outcomes.
Area of Science:
- Gastroenterology
- Medical Devices
- Animal Research
Background:
- Adequate bowel preparation is critical for colonoscopy quality, impacting procedure efficiency and diagnostic accuracy.
- Poor bowel preparation leads to increased procedure time, missed adenomas, higher costs, and decreased patient satisfaction.
- The Pure-Vu system is designed for intraprocedural colon cleansing, using irrigation and evacuation to manage inadequate prep.
Purpose of the Study:
- To evaluate the cleansing efficacy of the Pure-Vu system in animal models with inadequate bowel preparation.
- To assess the usability and safety of the Pure-Vu system during live animal colonoscopies.
Main Methods:
- The Pure-Vu system was utilized by four experienced gastroenterologists in 35 swine with intentionally reduced bowel preparation.
- Baseline bowel preparation was assessed, followed by colon cleansing with Pure-Vu, and subsequent reassessment.
- The system was attached to the colonoscope for intraprocedural use.
Main Results:
- No adverse events or procedure terminations were reported, indicating a favorable safety profile.
- Baseline adequate bowel preparation was achieved in 14% of swine, which improved to 100% after Pure-Vu use (p < .001).
- Physicians reported the Pure-Vu system as easy and intuitive to operate.
Conclusions:
- The Pure-Vu system demonstrated significant effectiveness in cleansing inadequately prepared swine colons.
- The device proved to be user-friendly and safe in the study.
- Clinical application of Pure-Vu may enhance overall colonoscopy prep quality and decrease the need for repeat procedures.
More Related Videos
15:49Flexible Colonoscopy in Mice to Evaluate the Severity of Colitis and Colorectal Tumors Using a Validated Endoscopic Scoring System
Published on: October 16, 2013
06:28E-Patient Counseling Trial E-PACO: Computer Based Education versus Nurse Counseling for Patients to Prepare for Colonoscopy
Published on: August 1, 2019
Related Concept Videos
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Endoscopic Procedures II: Colonoscopy
Endoscopic Procedures III: Video Capsule Endoscopy
Assessment of the Rectum and Anus
Rectal Inspection
Begin by inspecting the perianal and anal areas for color, texture, rashes,...
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
Lower GI Series: Barium Enema
Procedure Details
The examination begins by inserting a lubricated rectal tube into the patient's rectum to administer a radiopaque barium solution. The barium flow is carefully...