Balloon electric shock ablation. Effects on ventricular structure, function, and electrophysiology.
L L Mickleborough1, G J Wilson, L Harris
1Division of Cardiovascular Surgery, Toronto General Hospital, Ontario, Canada.
The Journal of Thoracic and Cardiovascular Surgery
|January 1, 1989
Summary
We developed balloon electric shock ablation to create scar tissue for treating ventricular tachycardia. This method effectively created scar tissue in animal hearts without significantly impacting overall heart function.
Area of Science:
- Cardiovascular Surgery
- Electrophysiology
- Medical Devices
Background:
- Transatrial balloon mapping is a new technique for intraoperative endocardial mapping of ventricular tachycardia.
- Balloon electric shock ablation aims to create homogenous scar tissue in the subendocardial region.
- Limited data exist on the effects of balloon electric shock ablation on myocardial tissue.
Purpose of the Study:
- To evaluate the effects of balloon electric shock ablation on ventricular function and myocardial tissue in an animal model.
- To assess the safety and efficacy of creating scar tissue using this novel ablation technique.
Main Methods:
- Balloon electric shock ablation was performed in animal models using nine cathodal shocks (100 J each) via a 4 cm² electrode grid.
- Ventricular function was assessed at 6 weeks using gated nuclear ventriculograms and pressure-volume relationships.
- Comparison was made with animals undergoing ventriculotomy and control animals.
Main Results:
- Balloon electric shock ablation led to decreased ventricular compliance, which was statistically significant compared to controls.
- Myocardial performance remained unchanged across all groups.
- Electrophysiologic studies failed to induce ventricular arrhythmias in the ablation group.
- Histological analysis revealed a homogenous, mature scar with defined borders, and no significant damage to surrounding myocardium or mitral valve.
Conclusions:
- Delivery of electric shocks via a balloon electrode grid can create homogenous, electrically inert scar tissue.
- Balloon electric shock ablation in healthy animal hearts did not significantly affect the structure and function of the left ventricle.
- This technique shows promise for creating targeted myocardial scars for arrhythmia treatment.