Reverse double inversion-recovery: Improving motion robustness of cardiac T2 -weighted dark-blood turbo spin-echo
Chenxi Hu1, Steffen Huber1,2, Syed R Latif2
1Department of Radiology and Biomedical Imaging, Yale School of Medicine, New Haven, Connecticut, USA.
Background:
Cardiac dark-blood turbo spin-echo (TSE) imaging is sensitive to through-plane motion, resulting in myocardial signal reduction.
Purpose:
To propose and validate reverse double inversion-recovery (RDIR)-a dark-blood preparation with improved motion robustness for the cardiac dark-blood TSE sequence.
Study Type:
Prospective.
Population:
Healthy volunteers (n = 10) and patients (n = 20).
Field Strength:
1.5T (healthy volunteers) and 3T (patients).
Assessment:
Compared to double inversion recovery (DIR), RDIR swaps the two inversion pulses in time and places the slice-selective 180° in late-diastole of the previous cardiac cycle to minimize slice misregistration. RDIR and DIR were performed in the same left-ventricular basal short-axis slice. Healthy subjects were imaged with two preparation slice thicknesses, 110% and 200%, while patients were imaged using a 200% slice thickness only. Images were assessed quantitatively, by measuring the myocardial signal heterogeneity and the extent of dropout, and also qualitatively on a 5-point scale.
Statistical Tests:
Quantitative and qualitative data were assessed with Student's t-test and Wilcoxon signed-rank test, respectively.
Results:
In healthy subjects, RDIR with 110% slice thickness significantly reduced signal heterogeneity in both the left ventricle (LV) and right ventricle (RV) (LV: P = 0.006, RV: P < 0.0001) and the extent of RV dropout (P < 0.0001), while RDIR with 200% slice thickness significantly reduced RV signal heterogeneity (P = 0.001) and the extent of RV dropout (P = 0.0002). In patients, RDIR significantly reduced RV myocardial signal heterogeneity (0.31 vs. 0.43; P = 0.003) and the extent of RV dropout (24% vs. 46%; P = 0.0005). LV signal heterogeneity exhibited a trend towards improvement with RDIR (0.12 vs. 0.16; P = 0.06). Qualitative evaluation showed a significant improvement of LV and RV visualization in RDIR compared to DIR (LV: P = 0.04, RV: P = 0.0007) and a significantly improved overall image quality (P = 0.03).
Data Conclusion:
RDIR TSE is less sensitive to through-plane motion, potentiating increased clinical utility for black-blood TSE.
Level Of Evidence:
1 Technical Efficacy: Stage 1 J. Magn. Reson. Imaging 2018;47:1498-1508.
More Related Videos
08:41Assessment of Cardiac Function and Myocardial Morphology Using Small Animal Look-locker Inversion Recovery SALLI MRI in Rats
Published on: July 19, 2013
06:29Cardiac Magnetic Resonance for the Evaluation of Suspected Cardiac Thrombus: Conventional and Emerging Techniques
Published on: June 11, 2019
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Magnetic Resonance Imaging
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...