Bore Disruption: An Unusual Mechanical Failure of Two Hip Hemiarthroplasties

Federico Dettoni1, Umberto Cottino1, Silvia Spriano2

  • 1Ospedale Mauriziano Umberto I, SCDU Ortopedia e Traumatologia, Turin, Italy.

Joints
|November 9, 2017
PubMed

We report a case of two hemiarthroplasties (HAs) that underwent an unexpected mechanical failure. Two patients affected by femoral fractures were treated with an HA. At 5 and 7 years after surgery, they showed a breakage of the HA at the head-stem junction. We analyzed macroscopically and microscopically (by a scanning electron microscopy) the surfaces of the broken prostheses. In both cases, a fracture was detected starting at the joining point between the femoral head and the slot that receives the taper cone of the stem (the so-called bore), with the typical macroscopic and microscopic pattern of a fatigue fracture. The fatigue fracture resulted from repeated cycles of torsional stresses throughout the years. In literature, we did not find any other report of such HA failure.