Delay Between Shaking and Actuation of a Hydrofluoroalkane Fluticasone Pressurized Metered-Dose Inhaler
Ariel Berlinski1, Dirk von Hollen2, John N Pritchard3
1Pulmonology Section, Department of Pediatrics, University of Arkansas for Medical Sciences, College of Medicine/Pediatric Aerosol Research Laboratory, Little Rock, Arkansas. berlinskiariel@uams.edu.
Background:
Inhaled corticosteroids are used to treat pediatric asthma. The shaking of a pressurized metered-dose inhaler (pMDI) is required to ensure consistency of emitted dose. Delays between shaking and actuating the pMDI are frequent during administration of aerosols to children where a valved holding chamber is used.
Methods:
In a recent clinical trial, we used a monitoring device to record shaking and actuation of the pMDI and the inhalation profiles of children with asthma while they were inhaling fluticasone hydrofluoroalkane from a valved holding chamber onto an external filter. During the procedure, in vitro and transport samples were generated without a delay between shaking and actuating the pMDI. Emitted dose, expressed as percentage of ex-actuator nominal dose, obtained from the second actuation following a recorded shake-actuation interval for subjects and from in vitro/transport samples (no delay) were compared.
Results:
The mean emitted dose was 158.6% (95% CI 150.1-167.2%) (subjects) and 106.8% (95% CI 104.7-108.9%) (in vitro + transport) of the ex-actuator nominal dose (P < .001). The mean delay between shaking and actuating the pMDI was 12.9 s (95% CI 11.9-13.9 s) for the subject samples. A strong correlation was observed between shake and actuation delay and the emitted dose of the second actuation following the delay (Spearman correlation coefficient = 0.61). A 10-, 20-, and 30-s delay resulted in an emitted dose of the second actuation following the delay of 147, 187, and 227% of the ex-actuator nominal dose, respectively.
Conclusions:
Delays between shaking and actuating a corticosteroid suspension pMDI resulted in an increase in the emitted dose of the second actuation following the delay. This can be a common occurrence when doses are administered by a caregiver to a patient via a holding chamber. This should be addressed by practitioners educating patients and parents on proper inhaler use. (ClinicalTrials.gov registration NCT01714063.).
More Related Videos
04:59Author Spotlight: Developing a Disposable Dosator for Preclinical Testing of Dry Powder Inhalers in Small Animal Models
Published on: August 18, 2023
07:28Dry Powder and Nebulized Aerosol Inhalation of Pharmaceuticals Delivered to Mice Using a Nose-only Exposure System
Published on: April 6, 2017
Related Concept Videos
Inhaled Medications
Asthma-IV: Diagnostic and Management
Clinical Assessment for Asthma:
This is the first step in diagnosing and managing asthma. It includes:
Modified-Release Drug Delivery Systems: Stimuli-Activated
Modified-Release Drug Delivery Systems: Drug Release Characteristics