Related Experiment Video
Updated: Feb 18, 2026

Treating Low Back Pain in Failed Back Surgery Patients with Multicolumn-lead Spinal Cord Stimulation
Published on: June 26, 2018
Fully Implantable Peripheral Nerve Stimulation for Hemiplegic Shoulder Pain: A Multi-Site Case Series With Two-Year
Richard D Wilson1,2,3, Maria E Bennett4, Vu Q C Nguyen5
1MetroHealth Rehabilitation Institute of Ohio, MetroHealth Medical Center, Cleveland, OH, USA.
Objective:
To explore the feasibility and safety of a single-lead, fully implantable peripheral nerve stimulation system for the treatment of chronic shoulder pain in stroke survivors.
Participants:
Participants with moderate to severe shoulder pain not responsive to conservative therapies for six months.
Methods:
During the trial phase, which included a blinded sham introductory period, a percutaneous single-lead peripheral nerve stimulation system was implanted to stimulate the axillary nerve of the affected shoulder. After a three-week successful trial, participants received an implantable pulse generator with an electrode placed to stimulate the axillary nerve of the affected shoulder. Outcomes included pain, pain interference, pain-free external rotation range of motion, quality of life, and safety. Participants were followed for 24 months.
Results:
Twenty-eight participants underwent trial stimulation and five participants received an implantable pulse generator. The participants who received the implantable generator experienced an improvement in pain severity (p = 0.0002). All five participants experienced a 50% or greater pain reduction at 6 and 12 months, and four experienced at least a 50% reduction at 24 months. There was an improvement in pain interference (p < 0.0001). There was an improvement in pain-free external ROM (p = 0.003). There were no serious adverse events related to the device or to the procedure.
Conclusions:
This case series demonstrates the safety and efficacy of a fully implantable axillary PNS system for chronic HSP. Participants experienced reduction in pain, reduction in pain interference, and improved pain-free external rotation ROM. There were no serious adverse events associated with the system or the procedure.