A Methodological Approach to Non-invasive Assessments of Vascular Function and Morphology
Published on: February 7, 2015
Comparison of Different Automated Office Blood Pressure Measurement Devices: Evidence of Nonequivalence and Clinical
Félix Rinfret1, Lyne Cloutier2, Robert Wistaff3
1Institut de recherches cliniques de Montréal (IRCM), Montréal, Québec, Canada; Département des sciences biomédicales, Université de Montréal, Montréal, Québec, Canada.
Background:
Automated office blood pressure (AOBP) measuring devices are increasingly recommended as preferred blood pressure (BP) diagnostic tools, but it is unclear how they compare and how clinical environments impact their performance.
Methods:
This prospective randomized factorial parallel 4-group study compared BP estimates by BpTRU (VSM MedTech, Vancouver, BC, Canada) and Omron HEM 907 (Omron Healthcare, Kyoto, Japan) devices in closed vs open areas. Patients diagnosed with hypertension were recruited during office visits. After baseline open-room AOBP measurement with the BpTRU, patients had a second BP measurement with either the BpTRU or HEM 907 in either open or closed areas. Absolute BP levels and differences between the first and second measurements were compared. Diagnostic performance was also assessed.
Results:
Two hundred fifty-eight patients were studied. Their mean age was 66.2 ± 12.0 years, and 62% were men. The mean of first AOBP estimates was 127.4/73.3 mm Hg. Analyses of subsequent measurements revealed no influence of open or closed areas on BP means and diagnostic performance. Conversely, the Omron HEM 907 exceeded BpTRU systolic BP measurements by 4.6 mm Hg (< 0.01) in closed areas and by 3.9 mm Hg (< 0.01) in open areas. The discrepancy between devices was amplified at lower BP levels.
Conclusions:
Although different areas did not influence BP estimates, the Omron HEM 907 significantly exceeded BpTRU measurements on average and especially at lower BP levels. These differences should be considered when interchanging devices and could have clinical decision impacts in a population of patients treated for hypertension. Our results support the constant use of only 1 device type in a given clinic.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Pre-Procedural Guidelines for Assessing Blood Pressure
Errors occurring during blood pressure monitoring
Several factors...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement