Monte Carlo simulation of the full energy peak efficiency of an HPGe detector
Waseem Khan1, Qingmin Zhang1, Chaohui He1
1Department of Nuclear Science and Technology, School of Energy and Power Engineering, Xi'an Jiaotong University, Xi'an 710049, China.
Abstract:
This paper presents a Monte Carlo method to obtain the full energy peak efficiency (FEPE) curve for a High Purity Germanium (HPGe) detector, as it is difficult and time-consuming to measure the FEPE curve experimentally. The Geant4 simulation toolkit was adopted to establish a detector model since detector specifications provided by the nominal manufacturer are usually insufficient to calculate the accurate efficiency of a detector. Several detector parameters were optimized. FEPE curves for a given HPGe detectors over the energy range of 59.50-1836keV were obtained and showed good agreements with those measured experimentally. FEPE dependences on detector parameters and source-detector distances were investigated. A best agreement with experimental result was achieved for a certain detector geometry and source-detector distance.
More Related Videos
09:00Indoor Experimental Assessment of the Efficiency and Irradiance Spot of the Achromatic Doublet on Glass ADG Fresnel Lens for Concentrating Photovoltaics
Published on: October 27, 2017
14:01Making Record-efficiency SnS Solar Cells by Thermal Evaporation and Atomic Layer Deposition
Published on: May 22, 2015
Related Concept Videos
Gas Chromatography: Types of Detectors-II
High-Performance Liquid Chromatography: Types of Detectors