Missed detection of significant positive and negative shifts in gentamicin assay: implications for routine laboratory
Gus Koerbin1,2, Jiakai Liu3, Alex Eigenstetter1
1New South Wales Health Pathology, Chatswood, NSW 2067, Australia.
Introduction:
A product recall was issued for the Roche/Hitachi Cobas Gentamicin II assays on 25th May 2016 in Australia, after a 15 - 20% positive analytical shift was discovered. Laboratories were advised to employ the Thermo Fisher Gentamicin assay as an alternative. Following the reintroduction of the revised assay on 12th September 2016, a second reagent recall was made on 20th March 2017 after the discovery of a 20% negative analytical shift due to erroneous instrument adjustment factor.
Materials And Methods:
The practices of an index laboratory were examined to determine how the analytical shifts evaded detection by routine internal quality control (IQC) and external quality assurance (EQA) systems. The ability of the patient result-based approaches, including moving average (MovAvg) and moving sum of outliers (MovSO) approaches in detecting these shifts were examined.
Results:
Internal quality control data of the index laboratory were acceptable prior to the product recall. The practice of adjusting IQC target following a change in assay method resulted in the missed negative shift when the revised Roche assay was reintroduced. While the EQA data of the Roche subgroup showed clear negative bias relative to other laboratory methods, the results were considered as possible 'matrix effect'. The MovAvg method detected the positive shift before the product recall. The MovSO did not detect the negative shift in the index laboratory but did so in another laboratory 5 days before the second product recall.
Conclusions:
There are gaps in current laboratory quality practices that leave room for analytical errors to evade detection.
More Related Videos
11:31Establishment of a High-throughput Setup for Screening Small Molecules That Modulate c-di-GMP Signaling in Pseudomonas aeruginosa
Published on: June 30, 2016
12:03Antimicrobial Synergy Testing by the Inkjet Printer-assisted Automated Checkerboard Array and the Manual Time-kill Method
Published on: April 18, 2019
Related Concept Videos
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Estimation of k and VD of Aminoglycosides