Related Experiment Video
Updated: Feb 17, 2026

Investigation into Deep Breathing through Measurement of Ventilatory Parameters and Observation of Breathing Patterns
Published on: September 16, 2019
Neural Breathing Pattern and Patient-Ventilator Interaction During Neurally Adjusted Ventilatory Assist and
Pradeep V Mally1, Jennifer Beck2,3,4, Christer Sinderby2,3,4,5
1Division of Neonatology, NYU School of Medicine, New York, NY.
Objective:
To compare neurally adjusted ventilatory assist and conventional ventilation on patient-ventilator interaction and neural breathing patterns, with a focus on central apnea in preterm infants.
Design:
Prospective, observational cross-over study of intubated and ventilated newborns. Data were collected while infants were successively ventilated with three different ventilator conditions (30 min each period): 1) synchronized intermittent mandatory ventilation (SIMV) combined with pressure support at the clinically prescribed, SIMV with baseline settings (SIMVBL), 2) neurally adjusted ventilatory assist, 3) same as SIMVBL, but with an adjustment of the inspiratory time of the mandatory breaths (SIMV with adjusted settings [SIMVADJ]) using feedback from the electrical activity of the diaphragm).
Setting:
Regional perinatal center neonatal ICU.
Patients:
Neonates admitted in the neonatal ICU requiring invasive mechanical ventilation.
Measurements And Main Results:
Twenty-three infants were studied, with median (range) gestational age at birth 27 weeks (24-41 wk), birth weight 780 g (490-3,610 g), and 7 days old (1-87 d old). Patient ventilator asynchrony, as quantified by the NeuroSync index, was lower during neurally adjusted ventilatory assist (18.3% ± 6.3%) compared with SIMVBL (46.5% ±11.7%; p < 0.05) and SIMVADJ (45.8% ± 9.4%; p < 0.05). There were no significant differences in neural breathing parameters, or vital signs, except for the end-expiratory electrical activity of the diaphragm, which was lower during neurally adjusted ventilatory assist. Central apnea, defined as a flat electrical activity of the diaphragm more than 5 seconds, was significantly reduced during neurally adjusted ventilatory assist compared with both SIMV periods. These results were comparable for term and preterm infants.
Conclusions:
Patient-ventilator interaction appears to be improved with neurally adjusted ventilatory assist. Analysis of the neural breathing pattern revealed a reduction in central apnea during neurally adjusted ventilatory assist use.
More Related Videos
05:28Electrophysiology on Isolated Brainstem-spinal Cord Preparations from Newborn Rodents Allows Neural Respiratory Network Output Recording
Published on: November 19, 2015
04:56Laryngeal Mask Airway LMA Placement in a Neonatal Patient Simulator Using a Non-Inflatable Supraglottic Airway SGA
Published on: July 14, 2023
Related Concept Videos
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation I: Indication and Settings
Neural Control of Respiration
Respiratory Centers in the Brainstem
Two primary areas comprise the respiratory center: the medullary respiratory center in the medulla oblongata and the pontine respiratory group in the pons. The...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Alterations in Respiration II
In Biot's breathing, the respiratory rate and depth are irregular, alternating between periods of deep gasping and apnea. Common causes...