Improvement of the repeatability of parallel transmission at 7T using interleaved acquisition in the calibration scan
Related Experiment Video
Updated: Feb 17, 2026

Author Spotlight: Introduction to Active Probe Atomic Force Microscopy with Quattro-Parallel Cantilever Arrays
Published on: June 13, 2023
Hiroyuki Kameda1,2, Kohsuke Kudo3,4, Tsuyoshi Matsuda2
1Department of Radiation Medicine, Hokkaido University Graduate School of Medicine, Sapporo, Japan.
Background:
Respiration-induced phase shift affects B0 /B1+ mapping repeatability in parallel transmission (pTx) calibration for 7T brain MRI, but is improved by breath-holding (BH). However, BH cannot be applied during long scans.
Purpose:
To examine whether interleaved acquisition during calibration scanning could improve pTx repeatability and image homogeneity.
Study Type:
Prospective.
Subjects:
Nine healthy subjects.
Field Strength/Sequence:
7T MRI with a two-channel RF transmission system was used.
Assessment:
Calibration scanning for B0 /B1+ mapping was performed under sequential acquisition/free-breathing (Seq-FB), Seq-BH, and interleaved acquisition/FB (Int-FB) conditions. The B0 map was calculated with two echo times, and the B1+ map was obtained using the Bloch-Siegert method. Actual flip-angle imaging (AFI) and gradient echo (GRE) imaging were performed using pTx and quadrature-Tx (qTx). All scans were acquired in five sessions. Repeatability was evaluated using intersession standard deviation (SD) or coefficient of variance (CV), and in-plane homogeneity was evaluated using in-plane CV.
Statistical Tests:
A paired t-test with Bonferroni correction for multiple comparisons was used.
Results:
The intersession CV/SDs for the B0 /B1+ maps were significantly smaller in Int-FB than in Seq-FB (Bonferroni-corrected P < 0.05 for all). The intersession CVs for the AFI and GRE images were also significantly smaller in Int-FB, Seq-BH, and qTx than in Seq-FB (Bonferroni-corrected P < 0.05 for all). The in-plane CVs for the AFI and GRE images in Seq-FB, Int-FB, and Seq-BH were significantly smaller than in qTx (Bonferroni-corrected P < 0.01 for all).
Data Conclusion:
Using interleaved acquisition during calibration scans of pTx for 7T brain MRI improved the repeatability of B0 /B1+ mapping, AFI, and GRE images, without BH.
Level Of Evidence:
1 Technical Efficacy Stage 1 J. Magn. Reson. Imaging 2017.
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Calibration Curves: Correlation Coefficient
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...