Discordance between mass spectrometry and immunometric IGF-1 assay in pituitary disease: a prospective study
Vivien Bonert1, John Carmichael2, Zengru Wu3
1Pituitary Center, Department of Medicine, Cedars-Sinai Medical Center, 127 S. San Vicente Blvd., Advanced Health Sciences Pavilion, Sixth Floor, A6600, Los Angeles, CA, 90048, USA. Vivien.Bonert@cshs.org.
Purpose:
Measuring IGF-1, a biomarker for GH activity, is critical to evaluating disordered hypothalamic-pituitary GH axis. Inconsistent IGF-1 measurements among different immunoassays are well documented. We switched from Immulite 2000 immunoassay to narrow-mass-extraction, high-resolution liquid chromatography mass-spectrometry (LC-MS) compliant with recent consensus recommendations on assay standardization. Comparability of these two assays in patients with pituitary disease in a clinical practice setting is not known. We sought to compare IGF-1 levels on Immulite 2000 and LC-MS in samples from naïve and treated patients with secretory and non-secretory pituitary masses.
Methods:
We prospectively collected serum samples from 101 patients treated at the Cedars-Sinai Pituitary Center between February 2012 and March 2014. We intentionally recruited more patients with acromegaly or GH deficiency to ensure a clinically representative cohort. Samples were classified as in or out of the respective reference ranges. Bland-Altman analysis was used to assess agreement between assays.
Results:
Twenty-four percent of samples were classified differently as below, in, or above range. Agreement between the assays was poor overall, with a significant bias for immunoassay reporting higher values than LC-MS. This pattern was also observed in patients with acromegaly and those with ≥ 2 pituitary hormone deficiencies.
Conclusions:
IGF-1 results may differ after switching from an older immunoassay to a consensus-compliant assay such as LC-MS. Clinicians should consider the potential impact of assay switching before altering treatment due to discrepant results, particularly in patients monitored over time, such as those with acromegaly and GH deficiency.