Automated QT analysis on Holter monitors in pediatric patients can differentiate long QT syndrome from controls
Christopher W Follansbee1, Lee Beerman1, Gaurav Arora1
1Division of Pediatric Cardiology, Children's Hospital of Pittsburgh of UPMC, Pittsburgh, PA, USA.
Background:
Borderline QTc is a common referral to the pediatric cardiology clinic. Evaluation is challenging due to significant overlap of normal and abnormal QTc ranges. We hypothesized that automated QT analysis on Holter could differentiate between patients with long QT syndrome (LQTS) and healthy controls.
Methods:
We conducted a retrospective review of 39 patients with known genotype-positive, phenotype-positive LQTS who underwent Holter monitoring between January 2010 and January 2016. They were compared 2:1 to age- and sex-matched controls. Automated QT analysis data were analyzed.
Results:
Significant differences were found in all automated QT and QTc fields, except minimum QTc interval (P = 0.57). Mean QTc interval (LQTS 479 ± 28 ms vs controls 429 ± 16 ms; P ≤ 0.001) and percent QTc intervals (%QTc) >450 ms (LQTS 80 ± 28% vs controls 14 ± 16%; P ≤ 0.001) were selected for further analysis. A receiver operating characteristic curve was generated for each variable demonstrating high area under the curve values of 0.9494 and 0.9540, respectively. Threshold values of ≥461 ms for mean QTc (sensitivity 79.49%, specificity 98.72%) and ≥65% of %QTc >450 ms (sensitivity 79.49%, specificity 98.72%) allowed highly specific discrimination between cohorts (false positive rate 1.09%). Similarly, thresholds of <434 ms (sensitivity 97.44, specificity 61.54) for mean QTc and <32% (sensitivity 89.74, specificity 87.18) for %QTc >450 ms resulted in highly sensitive discrimination (false negative rates 2.17% and 8.7%).
Conclusion:
Holter monitor testing with automated QT analysis may be a useful tool to differentiate LQTS and control patients.
More Related Videos
18:11A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
Published on: December 28, 2012
11:13Electrophysiological Analysis of human Pluripotent Stem Cell-derived Cardiomyocytes hPSC-CMs Using Multi-electrode Arrays MEAs
Published on: May 12, 2017
Related Concept Videos
Holter Monitor: 24-Hour Monitoring
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Dysrhythmias V: Evaluating Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias IV: Characteristics of Bradyarrhythmias