Comparison of a new Scheimpflug imaging combined with partial coherence interferometry biometer and a low-coherence
Ramón Ruiz-Mesa1, Antonio Abengózar-Vela1, María Ruiz-Santos1
1From the Oftalvist Centro Integral Ocular Jerez, Hospital Puerta del Sur, Jerez de la Frontera, Cádiz, Spain.
Purpose:
To evaluate the repeatability of a new biometer using Scheimpflug technology combined with partial coherence interferometry (PCI) (Pentacam AXL) and its agreement with a device based on optical low-coherence reflectometry (OLCR), the Allegro Biograph.
Setting:
Oftalvist Centro Integral Ocular Jerez, Jerez de la Frontera, Spain.
Design:
Evaluation of a diagnostic test.
Methods:
The mean keratometry (K), central corneal thickness (CCT), anterior chamber depth (ACD), and axial length (AL) were measured with the 2 devices 3 times by the same examiner in 2 groups (patients with cataract and patients without cataract). The repeatability was determined using the within-subject standard deviation, test-retest repeatability, coefficient of variation, and intraclass correlation coefficient. The correlation was evaluated with the Pearson coefficient and interchangeability with the Bland-Altman plot.
Results:
Eighty eyes (40 eyes in each group) of 80 patients were analyzed. Significant differences were found between the Scheimpflug-PCI device and the OLCR device for mean K in the normal group (P < .001) and for CCT in the normal group (P < .05) and the cataract group (P < .001). There were no differences between devices in ACD and AL in either group. The repeatability between devices was similar. Although a significant correlation between devices was found for all measurements (all P < .001), wide limits of agreement were found in both groups for all biometric parameters.
Conclusions:
The Scheimpflug-PCI and OLCR devices showed excellent intravisit repeatability and high correlation for mean K, CCT, ACD, and AL in healthy and cataractous eyes. No differences were found in AL; however, the 2 devices might not be interchangeable.
More Related Videos
11:57Three-dimensional Super Resolution Microscopy of F-actin Filaments by Interferometric PhotoActivated Localization Microscopy iPALM
Published on: December 1, 2016
10:39Measurement of X-ray Beam Coherence along Multiple Directions Using 2-D Checkerboard Phase Grating
Published on: October 11, 2016
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...