Related Experiment Video
Updated: Feb 17, 2026

Author Spotlight: Eco-Friendly Extraction of Bioactive Compounds Using Polyol-Based Microwave-Assisted Techniques
Published on: August 23, 2024
Solvent or thermal extraction of ethylene oxide from polymeric materials: Medical device considerations
Anne D Lucas1, Christopher Forrey1, David M Saylor1
1U.S. Food and Drug Administration, Center for Device and Radiological Health, Office of Science and Engineering Laboratories, Division of Biology, Chemistry and Materials Science, 10903 New Hampshire Ave, WO 64, Silver Spring, Maryland, 20903.
Abstract:
Ethylene oxide (EO) gas is commonly used to sterilize medical devices. Bioavailable residual EO, however, presents a significant toxicity risk to patients. Residual EO is assessed using international standards describing extraction conditions for different medical device applications. We examine a series of polymers and explore different extraction conditions to determine residual EO. Materials were sterilized with EO and exhaustively extracted in water, in one of three organic solvents, or in air using thermal desorption. The EO exhaustively extracted varies significantly and is dictated by two factors: the EO that permeates the material during sterilization; and the effectiveness of the extraction protocol in flushing residual EO from the material. Extracted EO is maximized by a close matches between Hildebrand solubility parameters δpolymer , δEO , and δsolvent . There remain complexities to resolve, however, because maximized EO uptake and detection are accompanied by great variability. These observations may inform protocols for material selection, sterilization, and EO extraction. © 2017 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 106B: 2455-2463, 2018.
Related Concept Videos
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Types of Step-Growth Polymers: Polyesters
Polyesters are commonly prepared from terephthalic acid and ethylene glycol; the crude product is known as poly(ethylene terephthalate) or PET. However, polyesters are synthesized industrially by transesterification of dimethyl terephthalate with ethylene glycol at 150 °C. The two reactants and the polymer...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...