Background:
Combined esophageal high-resolution impedance manometry (HRIM) measures multiple pressures and bolus transit simultaneously, facilitating detailed assessment of esophageal motility. Currently, normative values for water-perfused HRIM systems for Chinese populations are lacking.
Methods:
Healthy volunteers were enrolled for comprehensive anthropometric measures, blood biochemistry tests, and an HRIM study using 22 water-perfused pressure sensors and 12 impedance channels. Ten 5-mL liquid swallows of saline at 30-second intervals were conducted. The following parameters were calculated: distal contractile integral (DCI), distal latency (DL), lower esophageal sphincter (LES) basal pressure, 4-second integrated relaxation pressure (IRP-4s), and complete bolus transit percentage. Normal values were established based on the 5th and 95th percentiles.
Key Results:
All 66 participants (34 male, 32 female, aged 21-64 years) completed the study and tolerated the HRIM procedure well. The upper normal limit (95th percentile) of IRP-4 second was 20 mmHg. The 5th-95th percentile range for DCI, DL, and complete bolus transit was 99-2186 mmHg●s●cm, 6.2-11.3 second, and 50%-100%, respectively. Age was negatively correlated with DL. Females had significantly higher upper limits for IRP-4s and median DCI than males. Multivariate analyses confirmed that IRP-4s was higher in females, and that higher body mass index and waist circumference were associated with reduced DL and better bolus transit, respectively.
Conclusions And Inferences:
We established normative values for the water-perfused HRIM system for a Chinese population. Gender and anthropometric factors may affect various major HRIM parameters and should be taken into account when interpreting HRIM results in clinical practice.