Near-Field Ultrasound Imaging During Radiofrequency Catheter Ablation: Tissue Thickness and Epicardial Wall
David E Haines1, Matthew Wright2, Erik Harks2
1From the Department of Cardiovascular Medicine, Beaumont Health System and Oakland University William Beaumont School of Medicine, Royal Oak, MI (D.E.H.,); St. Thomas' Hospital, London, United Kingdom (M.W.); Philips Healthcare, Best, The Netherlands (E.H., S.D., S.F., R.B.); Philips Research, Eindhoven, The Netherlands (H.B., A.F.K., N.M., F.Z.); and Boston Scientific Co. Inc, San Jose, CA (D.R., W.S., D.C., J.C.). dhaines@beaumont.edu.
Background:
Safe and successful radiofrequency catheter ablation depends on creation of transmural lesions without collateral injury to contiguous structures. Near-field ultrasound (NFUS) imaging through transducers in the tip of an ablation catheter may provide important information about catheter contact, wall thickness, and ablation lesion formation.
Methods And Results:
NFUS imaging was performed using a specially designed open-irrigated radiofrequency ablation catheter incorporating 4 ultrasound transducers. Tissue/phantom thickness was measured in vitro with varying contact angles. In vivo testing was performed in 19 dogs with NFUS catheters positioned in 4 chambers. Wall thickness measurements were made at 222 sites (excluding the left ventricle) and compared with measurements from intracardiac echocardiography. Imaging was used to identify the epicardium with saline infusion into the pericardial space at 39 sites. In vitro, the measured exceeded actual tissue/phantom thickness by 13% to 20%. In vivo, NFUS reliably visualized electrode-tissue contact, but sensitivity of epicardial imaging was 92%. The chamber wall thickness measured by NFUS correlated well with intracardiac echocardiography (r=0.86; P<0.0001). Sensitivity of lesion identification by NFUS was 94% for atrial and 95% for ventricular ablations. NFUS was the best parameter to predict lesion depth in right and left ventricle (r=0.47; P<0.0001; multiple regression P=0.0025). Lesion transmurality was correctly identified in 87% of atrial lesions.
Conclusions:
NFUS catheter imaging reliably assesses electrode-tissue contact and wall thickness. Its use during radiofrequency catheter ablation may allow the operator to assess the depth of ablation required for transmural lesion formation to optimize power delivery.
More Related Videos
28:13Catheter Ablation in Combination With Left Atrial Appendage Closure for Atrial Fibrillation
Published on: February 26, 2013
07:38Real-time Monitoring of High Intensity Focused Ultrasound HIFU Ablation of In Vitro Canine Livers Using Harmonic Motion Imaging for Focused Ultrasound HMIFU
Published on: November 3, 2015
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...