Related Experiment Video
Updated: Feb 16, 2026

Controllable Nucleation of Cavitation from Plasmonic Gold Nanoparticles for Enhancing High Intensity Focused Ultrasound Applications
Published on: October 5, 2018
Quantitative imaging performance of frequency-tunable capacitive micromachined ultrasonic transducer array designed
Martin Pekař1, Nenad Mihajlović2, Harm Belt2
1Philips Research, Royal Philips NV, Eindhoven, The Netherlands; Thorax Center Dept. of Biomedical Engineering, Erasmus MC, Rotterdam, The Netherlands.
Abstract:
Commercially available intracardiac echo (ICE) catheters face a trade-off between viewing depth and resolution. Frequency-tunable ICE probes would offer versatility of choice between penetration or resolution imaging within a single device. In this phantom study, the imaging performance of a novel, frequency-tunable, 32-element, 1-D CMUT array integrated with front-end electronics is evaluated. Phased-array ultrasound imaging with a forward-looking CMUT probe prototype operated beyond collapse mode at voltages up to three times higher than the collapse voltage (-65 V) is demonstrated. Imaging performance as a function of bias voltage (-70 V to -160 V), transmit pulse frequency (5-25 MHz), and number of transmit pulse cycles (1-3) is quantified, based on which penetration, resolution, and generic imaging modes are identified. It is shown that by utilizing the concept of frequency tuning, images with different characteristics can be generated trading-off the resolution and penetration depth. The penetration mode provides imaging up to 71 mm in the tissue-mimicking phantom, axial resolution of 0.44 mm, and lateral resolution of 0.12 rad. In the resolution mode, axial resolution of 0.055 mm, lateral resolution of 0.035 rad, and penetration depth of 16 mm are measured. These results show what this CMUT array has the potential versatile characteristics needed for intracardiac imaging, despite its relatively small transducer aperture size of 2 mm × 2 mm imposed by the clinical application.
More Related Videos
16:01An Experimental Protocol for Assessing the Performance of New Ultrasound Probes Based on CMUT Technology in Application to Brain Imaging
Published on: September 24, 2017
08:08Evaluating Targeting Accuracy in the Focal Plane for an Ultrasound-guided High-intensity Focused Ultrasound Phased-array System
Published on: March 6, 2019