Related Experiment Video
Updated: Feb 16, 2026

The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
Published on: August 4, 2022
Fracture at the Modular Junction of a Cementless Revision Hip System: A Case Report
Stephen T Duncan1, Christopher B Hayes1, Ryan M Nunley2
1Department of Orthopaedic Surgery, Kentucky Clinic, University of Kentucky, Lexington, Kentucky.
Case:
Multiple cases of modular femoral component failure involving fatigue at the modular stem junction have been reported following revision total hip arthroplasty. In the present report, we describe what we believe to be the first case of catastrophic failure involving the Biomet Arcos revision system. The failure involved a young patient with an elevated body mass index and proximal bone deficiency.
Conclusion:
The possibility of this failure mechanism must be weighed against the potential advantages of modular cementless stems when planning for complex hip arthroplasty, especially in the young, active patient.