Related Experiment Video
Updated: Feb 16, 2026

The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
Published on: August 4, 2022
Intraprosthetic Dislocation of Dual-Mobility Bearings Associated with Closed Reduction: A Report of Two Cases
Gregory L Cvetanovich1, Yale A Fillingham1, Craig J Della Valle1
1Midwest Orthopaedics at Rush, 1611 West Harrison Avenue, Chicago, IL 60612. E-mail address for G.L. Cvetanovich: gregory.cvetanovich@gmail.com.
Case:
We present two patients with dual-mobility total hip arthroplasty components who underwent closed reduction of posterior dislocations with conscious sedation in the emergency room. Following closed reduction, both patients had immediate pain, clicking, and grinding of the hip. Radiographs identified intraprosthetic dislocation of the dual-mobility components, and revision surgery was required.
Conclusion:
When dislocation of a dual-mobility bearing occurs, the surgeon should consider performing a careful closed reduction with muscle paralysis and use of fluoroscopic guidance in the operating room to avoid intraprosthetic dislocation.
More Related Videos
04:41Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
05:57A Teleoperated Robotic System-Assisted Percutaneous Transiliac-Transsacral Screw Fixation Technique
Published on: January 6, 2023