Toward Calibration-Free Continuous Glucose Monitoring Sensors: Bayesian Calibration Approach Applied to
Giada Acciaroli1, Martina Vettoretti1, Andrea Facchinetti1
1Department of Information Engineering, University of Padova , Padova, Italy .
Background:
Continuous glucose monitoring (CGM) sensors need to be calibrated twice/day by using self-monitoring of blood glucose (SMBG) samples. Recently, to reduce the calibration frequency, we developed an online calibration algorithm based on a multiple-day model of sensor time variability and Bayesian parameter estimation. When applied to Dexcom G4 Platinum (DG4P) sensor data, the algorithm allowed the frequency of calibrations to be reduced to one-every-four-days without significant worsening of sensor accuracy. The aim of this study is to assess the same methodology on raw CGM data acquired by a next-generation Dexcom CGM sensor prototype and compare the results with that obtained on DG4P.
Methods:
The database consists of 55 diabetic subjects monitored for 10 days by a next-generation Dexcom CGM sensor prototype. The new calibration algorithm is assessed, retrospectively, by simulating an online procedure using progressively fewer SMBG samples until zero. Accuracy is evaluated with mean absolute relative differences (MARD) between blood glucose versus CGM values.
Results:
The one-per-day and one-every-two-days calibration scenarios in the next-generation CGM data have an accuracy of 8.5% MARD (vs. 11.59% of DG4P) and 8.4% MARD (vs. 11.63% of DG4P), respectively. Accuracy slightly worsens to 9.2% (vs. 11.62% of DG4P) when calibrations are reduced to one-every-four-days. The calibration-free scenario results in 9.3% MARD (vs. 12.97% of DG4P).
Conclusions:
In next-generation Dexcom CGM sensor data, the use of an online calibration algorithm based on a multiple-day model of sensor time variability and Bayesian parameter estimation aids in the shift toward a calibration-free scenario with even better results than those obtained in present-generation sensors.