Related Experiment Video
Updated: Feb 16, 2026

Echocardiographic Evaluation of Atrial Communications before Transcatheter Closure
Published on: February 8, 2022
Usefulness of Sealants for Dural Closure: Evaluation in an In Vitro Model
Tristan van Doormaal1,2, Ahmet Kinaci1, Sander van Thoor2
1Department of Neurosurgery, Brain Center Rudolph Magnus, University Medical Centre Utrecht, Utrecht, The Netherlands.
Background:
Cerebrospinal fluid (CSF) leakage occurs in 4% to 32% of cranial surgeries and is associated with significant patient burden and expense. The use of sealant as an adjunct to primary dural closure is assumed to help prevent CSF leakage.
Objective:
To examine the utility of different sealants for dural closure using an in Vitro model.
Methods:
We evaluated 9 commonly used dural sealants, including Tachosil (Takeda Inc, Osaka, Japan), Adherus (Hyperbranch Inc, Durham, North Carolina), Duraform (Codman, Raynham, Massachusetts), Tissudura (Baxter, Deerfield, Illinois), Hemopatch (Baxter), TissuePatchDural (Tissuemed, Leeds, United Kingdom), Tisseel (Baxter), Duragen Secure (Integra, Plainsboro, New Jersey), and Duraseal, (Integra). Sealants were tested in 2 novel in Vitro setups using fresh porcine dura: the first tested the acute burst pressure of a sealed 3-mm gap, while the second examined resistance to a pressure wave mimicking intracranial pressure for 72 h.
Results:
Adherus showed the highest mean burst pressure (87 ± 47 mmHg) followed by Tachosil (71 ± 16 mmHg) and Duraseal (51 ± 42 mmHg); these were the only 3 sealants showing burst pressures above normal physiological intracranial pressure. In the 72-h setup, only Adherus and Duraseal maintained appropriate sealing for the duration of the experiment. Tachosil released from the dura after 1.4 h (95% confidence interval, -1.8-4.7).
Conclusion:
Given the high cost of sealants and the results of this study, we advocate a critical attitude toward sealant application as an adjunct to classic dural closure.