Measurement of absolute energy spectra from a clinical CT machine under working conditions using a Compton
1Department of Radiation Physics, Faculty of Health Sciences, University of Linköping, Sweden.
Abstract:
Absolute measurements of photon energy spectra (keV-1 sr-1 mA-1 s-1) from a clinical CT machine have been performed under normal working conditions (140 mA tube current) using a Compton spectrometer. The inaccuracy of the measured spectra is estimated to be +/- 6%, determined by uncertainties in dead-time corrections and in the parameters of the geometrical set-up. Absorbed doses measured in thermoluminescent LiF dosimeters agree within this uncertainty with calculated ones derived from measured spectra (80 kVp, 120 kVp and 140 kVp) and tabulated mass energy absorption coefficients for LiF. Comparison with tabulated energy spectra from the literature clearly shows the effect of the extremely small anode angle (7 degrees) in the CT machine (15 degrees and 17 degrees for the tabulated energy spectra).
More Related Videos
10:42In Depth Analyses of LEDs by a Combination of X-ray Computed Tomography (CT) and Light Microscopy (LM) Correlated with Scanning Electron Microscopy (SEM)
Published on: June 16, 2016
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Related Concept Videos
Spectrophotometry: Introduction
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Emission Spectroscopy: Overview
Atomic Emission Spectroscopy: Instrumentation
Atomic Emission Spectroscopy: Lab