Hybrid setup for micro- and nano-computed tomography in the hard X-ray range
Christian Fella1, Andreas Balles1, Randolf Hanke1
1Chair for X-ray Microscopy, Julius-Maximilians-Universität Würzburg, 97074 Würzburg, Germany.
A new compact X-ray microscope offers versatile non-destructive testing. This hybrid instrument combines micro-computed tomography (micro-CT) and high-resolution microscopy for detailed 3D sample analysis.
Area of Science:
- Materials Science
- Physics
- Engineering
Background:
- Non-destructive testing (NDT) is crucial for miniaturized industrial and medical applications.
- X-ray microscopy provides 3D volumetric data through computed tomography (CT).
Purpose of the Study:
- To describe a novel compact X-ray microscope for advanced NDT.
- To highlight its hybrid capabilities for multi-modal analysis.
Main Methods:
- Development of a compact hard X-ray microscope (9 keV) utilizing a liquid-metal-jet source.
- Implementation of two distinct operational modes: micro-CT without optics and magnified X-ray microscopy with optics.
- Enabling mode switching without sample repositioning.
Main Results:
- Micro-CT mode achieves 1 μm resolution for millimeter-sized samples.
- Microscopy mode resolves features down to 150 nm.
- The instrument integrates both modes seamlessly for comprehensive sample investigation.
Conclusions:
- The developed instrument advances laboratory-based multi-mode X-ray microscopy.
- It offers a significant step towards establishing high-resolution X-ray microscopy as a standard investigation method for microscopic samples.
More Related Videos
07:013D Imaging of Soft-Tissue Samples using an X-ray Specific Staining Method and Nanoscopic Computed Tomography
Published on: October 24, 2019
07:06Use of Micro X-ray Computed Tomography with Phosphotungstic Acid Preparation to Visualize Human Fibromuscular Tissue
Published on: September 5, 2019
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Range
15.9; 16.1; 15.2; 14.8; 15.8; 15.9; 16.0; 15.5
Measurements of the amount of soda in a 16-ounce can vary since different subjects record these measurements or since the exact amount - 16 ounces of liquid, was not...
Hybrid Zones
Variation: Normal Distribution, Range, and Standard Deviation