Optical Measurements during Deep Brain Stimulation Lead Implantation: Safety Aspects
Peter Zsigmond1, Simone Hemm-Ode, Karin Wårdell
1Department of Neurosurgery and Clinical and Experimental Medicine, Linköping University, Linköping, Sweden.
Background:
Intracerebral hemorrhage (ICH) is the most feared complication in deep brain stimulation (DBS) surgery. The aim of the study was to evaluate patient safety and outcome using laser Doppler flowmetry (LDF) as guidance tool during DBS implantations.
Methods:
An LDF probe adapted for the stereotactic system was used as guide for creation of the trajectory. The microcirculation along 83 preplanned trajectories was measured with the guide during DBS surgery for movement disorders. The microvascular blood flow levels were investigated for all measurement positions. Medical record and postoperative radiology were retrospectively reviewed.
Results:
Of 2,963 measurement positions, 234 (7.9%) showed at least a doubled blood flow compared to the surrounding tissue. Of these 2.2% had a more than 5 times higher blood flow in front of the probe tip. Along 1 trajectory, a small ICH was detected during surgery. Increased blood flow was more common close to sulci and verticals.
Conclusion:
Real-time LDF measurement of the microcirculation using a forward-looking probe during DBS surgery can detect blood flow peaks and further minimize the risk of developing ICH. No separate guide tube is necessary as the probe also creates the trajectory for the DBS lead.
Related Concept Videos
Survey Safety
Household Wiring And Electrical Safety
Phase-lead and Phase-lag Controllers
Quantitative Aspects of Drug-Receptor Interaction
Stimulants
Cocaine can be administered via snorting, injection, or smoking. It primarily functions by blocking the reuptake of dopamine, resulting in a euphoric high characterized by an intense sensation of happiness and...
Time and frequency -Domain Interpretation of Phase-lead Control
The design of phase-lead control involves the strategic placement of poles and zeros to balance steady-state error and system...