Related Experiment Video
Updated: Jul 23, 2026

10:05
Anatomically Realistic Neonatal Heart Model for Use in Neonatal Patient Simulators
Published on: February 5, 2019
Technical realization of a sensorized neonatal intubation skill trainer for operators' retraining and a pilot study
Davide Panizza1, Rosa T Scaramuzzo2,3, Francesca Moscuzza2,3
1Istituto di Scienze della Vita, Scuola Superiore Sant'Anna, Piazza Martiri della Libertà, 33, 56127, Pisa, Italy. dav.panizza@gmail.com.
Italian Journal of Pediatrics
|January 6, 2018
Summary
This study developed an active neonatal intubation trainer that provides objective feedback, significantly reducing excessive pressure during laryngoscopy. This innovative tool enhances neonatal intubation training and minimizes patient trauma risk.
Area of Science:
- Medical Simulation
- Pediatric Anesthesiology
- Neonatal Care
Background:
- Excessive pressure during neonatal endotracheal intubation can cause permanent injury.
- Existing low-fidelity trainers lack valid feedback mechanisms for tissue pressure.
- This study introduces an active neonatal intubation skill trainer with objective feedback capabilities.
Purpose of the Study:
- To technically realize and validate an active neonatal intubation skill trainer.
- To provide objective, real-time feedback on applied pressure during simulated neonatal intubation.
- To assess the trainer's effectiveness in improving procedural technique and reducing trauma risk.
Main Methods:
- Commercial force sensors were integrated into a Laerdal Neonatal Intubation Trainer, simulating a difficult intubation (Cormack and Lehane grade 3).
- An associated software provided real-time auditory feedback when pressure exceeded predefined thresholds.
- Participants underwent two intubation sessions with a 24-hour interval, including debriefing and interviews to gather performance and design feedback.
Main Results:
- Statistically significant learning trends were observed, with pressure on critical points (epiglottis and dental arches) decreasing significantly between sessions (p < 0.0001).
- The epiglottis sensor registered the highest stress.
- A significant inverse correlation was found between attempt duration and applied pressure, particularly in the second session (rS = 0.603).
Conclusions:
- The developed skill trainer serves as a reliable model for difficult neonatal intubation scenarios.
- The trainer has the potential to optimize procedures by controlling trauma risk.
- It offers a valuable tool for improving healthcare personnel retraining in neonatal intubation.