Related Experiment Video
Updated: Feb 16, 2026

Integrated Photoacoustic Ophthalmoscopy and Spectral-domain Optical Coherence Tomography
Published on: January 15, 2013
Correlation between frequency-domain optical coherence tomography and fractional flow reserve in
Francesco Burzotta1, Roberto Nerla1, Jonathan Hill2
1Institute of Cardiology, Catholic University of the Sacred Heart, Policlinico A. Gemelli, 8, 00168 Rome, Italy.
Background:
The decision-making process of patients with angiographically-intermediate coronary lesions (ICL) is clinically challenging and may benefit from adjunctive invasive techniques. Fractional-flow-reserve (FFR) represents the gold standard to evaluate ICL but frequency-domain optical-coherence-tomography (OCT) is a novel, promising, high resolution coronary imaging technique, which allows physiopathologic assessment of coronary plaque. We investigated the possible relation between OCT and FFR in selected ICL patients.
Methods:
Stable or unstable patients with ICL who underwent both FFR and OCT assessment at two large tertiary centers were retrospectively enrolled. FFR was performed according to standard methodology. OCT images were (on blind to clinical and FFR results) analyzed to assess minimal lumen area (MLA), percentage area stenosis (AS), thrombus and plaque ulceration.
Results:
Forty patients were identified (62±10years, 93% symptomatic, 35% acute presentation, 93% left-anterior-descending artery ICL). Percentage diameter stenosis at quantitative coronary angiography was 40±12% and FFR was 0.85±0.07. MLA (p=0.009), AS (p<0.001) and plaque ulceration (p=0.02) were significantly associated with FFR values. An integrated assessment of AS (≥ or <70%), MLA (≥ or <2.5mm2) and presence or absence of thrombus and plaque ulceration was found to have the potential to accurately (sensitivity 91%, specificity 93%) predict FFR results.
Conclusion:
In patients with ICL, a combination of different OCT parameters may help predict FFR results. These findings suggest that only a comprehensive assessment of lesion features by OCT can allow an accurate prediction of lesion severity assessed by FFR.
Related Concept Videos
Linear Approximation in Frequency Domain
In contrast, nonlinear systems do not inherently possess these properties. However, for small deviations around an operating point, a nonlinear system can often be approximated as linear....
Frequency-Domain Interpretation of PD Control
The proportional control gain, combined with the...
Time and frequency -Domain Interpretation of PI Control
Acting as a low-pass filter, the PI controller slows the system's response and extends settling times. This requires...
Time and frequency -Domain Interpretation of Phase-lead Control
The design of phase-lead control involves the strategic placement of poles and zeros to balance steady-state error and system...
Time and frequency -Domain Interpretation of Phase-lag Control
Phase-lag controllers do not place a pole at zero, but instead influence the steady-state error by amplifying any...
Correlations