Quantitative assessment of nonsolid pulmonary nodule volume with computed tomography in a phantom study
Marios A Gavrielides1, Benjamin P Berman2, Mark Supanich3
1Division of Imaging, Diagnostics, and Software Reliability, Office of Science and Engineering Laboratories, , Office of In Vitro Diagnostics and Radiological Health, Center for Devices and Radiological Health, U.S. Food and Drug Administration, Silver Spring, Maryland, USA.
Background:
To assess the volumetric measurement of small (≤1 cm) nonsolid nodules with computed tomography (CT), focusing on the interaction of state of the art iterative reconstruction (IR) methods and dose with nodule densities, sizes, and shapes.
Methods:
Twelve synthetic nodules [5 and 10 mm in diameter, densities of -800, -630 and -10 Hounsfield units (HU), spherical and spiculated shapes] were scanned within an anthropomorphic phantom. Dose [computed tomography scan dose index (CTDIvol)] ranged from standard (4.1 mGy) to below screening levels (0.3 mGy). Data was reconstructed using filtered back-projection and two state-of-the-art IR methods (adaptive and model-based). Measurements were extracted with a previously validated matched filter-based estimator. Analysis of accuracy and precision was based on evaluation of percent bias (PB) and the repeatability coefficient (RC) respectively.
Results:
Density had the most important effect on measurement error followed by the interaction of density with nodule size. The nonsolid -630 HU nodules had high accuracy and precision at levels comparable to solid (-10 HU) nonsolid, regardless of reconstruction method and with CTDIvol as low as 0.6 mGy. PB was <5% and <11% for the 10- and 5-mm in nominal diameter -630 HU nodules respectively, and RC was <5% and <12% for the same nodules. For nonsolid -800 HU nodules, PB increased to <11% and <30% for the 10- and 5-mm nodules respectively, whereas RC increased slightly overall but varied widely across dose and reconstruction algorithms for the 5-mm nodules. Model-based IR improved measurement accuracy for the 5-mm, low-density (-800, -630 HU) nodules. For other nodules the effect of reconstruction method was small. Dose did not affect volumetric accuracy and only affected slightly the precision of 5-mm nonsolid nodules.
Conclusions:
Reasonable values of both accuracy and precision were achieved for volumetric measurements of all 10-mm nonsolid nodules, and for the 5-mm nodules with -630 HU or higher density, when derived from scans acquired with below screening dose levels as low as 0.6 mGy and regardless of reconstruction algorithm.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Chronic Obstructive Pulmonary Disease-IV: Assessement and Diagnostic Studies
Medical History
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Pulmonary Embolism II: Diagnostic Studies and Interprofessional Care
Constant Volume Calorimetry