Precision of the Connection Between Implant and Standard or Computer-Aided Design/Computer-Aided Manufacturing
Purpose:
The aim of this in vitro study was to verify whether or not stock and computer-aided design/computer-aided manufacturing (CAD/CAM) abutments show similar precision in the connection with the respective implants.
Materials And Methods:
Ten CAD/CAM titanium abutments were compared with 10 stock titanium abutments. Each abutment fit a regular-platform implant (Institute Straumann). Implants and abutments were measured independently and then connected. During the connection procedure, the torque was measured using a six-axes load cell. Then, outer geometric features of the implant-abutment connection were measured again. Finally, the assembly was sectioned to provide the analysis of inner surfaces in contact. The geometric measurements were performed using a multisensored opto-mechanical coordinate measuring machine. The following parameters were measured and compared for the CAD/CAM and stock titanium abutment groups, respectively: width of interference and interference length between the conical surfaces of the implant and abutment; and volume of material involved in the implant-abutment connection.
Results:
Interference width mean ± SD values of 18 ± 0.5 and 14 μ 0.5 μm were calculated for the stock and CAD/CAM titanium abutment groups, respectively. The difference was statistically significant (P = .02). Furthermore, the interference length mean ± SD values of 763 ± 10 and 816 ± 43 μm were calculated for stock and CAD/CAM titanium abutment groups, respectively. The difference was also statistically significant (P = .04). Finally, the volume of material involved in the implant-abutment connection was compared between stock and CAD/CAM titanium abutment groups; the mean ± SD values of 0.134 ± 0.014 and 0.108 ± 0.023 mm3 were significantly different (P = .009).
Conclusion:
Both standard and CAD/CAM abutment groups showed a three-dimensional (3D) seal activation after the screw tightening. Nevertheless, stock titanium abutments showed a significantly higher volume of material involved in the implant-abutment connection compared with that of CAD/CAM titanium abutments.
More Related Videos
07:42Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
03:55Computer-Aided Three-Dimensional Visualization in the Treatment of Locally Advanced Thyroid Cancer
Published on: June 9, 2023
Related Concept Videos
Empirical Method to Interpret Standard Deviation
This rule is used widely in statistics to calculate the proportion of data values...
Steel Manufacturing
During this smelting process, limestone plays a crucial role by forming slag. Slag captures impurities within the molten iron, such...
Group Design
Standard Electrode Potentials
Uncertainty in Measurement: Accuracy and Precision
Dietary Connections