Quantitative method for analysis of six anticoagulant rodenticides in faeces, applied in a case with repeated samples
Kristin Opdal Seljetun1,2, Elin Eliassen3, Ritva Karinen3
1Department of Companion Animal Clinical Sciences, Faculty of Veterinary Medicine, Norwegian University of Life Sciences (NMBU), P.O. Box 8146 Dep, 0033, Oslo, Norway. kristin.opdal.seljetun@nmbu.no.
Background:
Accidental poisoning with anticoagulant rodenticides is not uncommon in dogs, but few reports of the elimination kinetics and half-lives in this species have been published. Our objectives were to develop and validate a new method for the quantification of anticoagulant rodenticides in canine blood and faeces using reversed phase ultra-high performance liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS) and apply the method on a case of anticoagulant rodenticide intoxication.
Results:
Sample preparation was liquid-liquid extraction. Six anticoagulant rodenticides were separated using a UPLC® BEH C18-column with a mobile phase consisting of 5 mM ammonium formate buffer pH 10.2 and methanol. MS/MS detection was performed with positive electrospray ionization and two multiple reaction monitoring transitions. The limits of quantification were set at the levels of the lowest calibrator (1.5-2.7 ng/mL or ng/g). The method was successfully applied to a case from a dog accidentally poisoned with anticoagulant rodenticide. Coumatetralyl and brodifacoum concentrations were determined from serial blood and faecal samples. A terminal half-life of at least 81 days for coumatetralyl in blood was estimated, which is longer than previous reported in other species. A slow elimination of brodifacoum from the faeces was found, with traces still detectable in the faeces at day 513.
Conclusions:
This study offers a new method of detection and quantification of six frequently used anticoagulant rodenticides in canine faeces. Such drugs might cause serious health effects and it is important to be able to detect these drugs, to initiate proper treatment. The very long elimination half-lives detected in our study is important to be aware of in assessment of anticoagulant rodenticide burden to the environment.
More Related Videos
Related Concept Videos
Anticoagulant Drugs: Vitamin K Antagonists and Direct Oral Anticoagulants
Warfarin, a prominent vitamin K antagonist family member, exerts its effect by inhibiting the enzyme VKORC1 (vitamin K epoxide reductase complex 1). By hindering this enzyme, warfarin...
Feces Formation and Defecation
The mass peristalsis then pushes the feces into the rectum, which stretches the rectal walls to activate...
Quantitative Analysis
In quantitative analysis, two key measurements are made: the sample quantity and a property proportional to the amount of the analyte (the substance being analyzed). This forms the basis of the...
Sampling Methods: Sample Types
Solid samples include a variety of substances, such as sediments from water bodies, soil, metals, and biological tissues. Two standard methods for extracting sediments from water bodies are grab sampling and piston coring. Grab sampling involves using a device to collect a discrete sediment sample from the bottom of a water body with minimal disturbance. Grab samples do not always represent the entire area due to...
Convenience Sampling Method
Convenience sampling is a non-random method of sample selection; this method selects individuals that are easily accessible and may result in biased data. For example, a marketing...
Sampling Methods: Overview
In analytical chemistry, the choice of...