Preliminary Evaluation of the Viability of Peritoneal Drainage Catheters Implanted in Rats for Extended Durations
Nathan Legband1, Arielle Black2, Craig Kreikemeier-Bower3
1a Department of Mechanical and Material Engineering , University of Nebraska-Lincoln , Lincoln , Nebraska , USA.
Abstract:
Purpose/Aim: In developing a novel peritoneal oxygenation therapy, catheters implanted into the peritoneal cavity became obstructed with omental tissue and prevented the infusion and removal of fluid from the peritoneal cavity. The obstruction of peritoneal catheters is a significant failure in researching various peritoneal treatments as further fluid administration is no longer possible. The purpose of this preliminary study was to determine the most effective catheter design for infusion and removal of fluid into the peritoneal cavity of rats. Materials and Methods: Four types of catheters were tested including the Jackson-Pratt, round fluted drain, flat fluted drain, and an original design. Three of each catheter type were surgically placed into the peritoneal cavity of rats (n = 12). In order to test the efficacy of each catheter, saline was infused and extracted twice daily. Catheters were scored on a weighted scale based on the amount of time they remained patent, the subjective force needed for extraction/infusion, and the amount of saline removed. Results: The round and flat fluted drain catheters remained patent for the full duration of the study (12 days) compared to the other models which failed after 7 days. These catheters also yielded a high average for extracted saline volume and an easy extraction/infusion. Conclusions: The round and flat fluted drain catheters were recognized as viable options to be used in rats for peritoneal drain studies of up to 12 days.
More Related Videos
10:07Cystometric and External Urethral Sphincter Measurements in Awake Rats with Implanted Catheter and Electrodes Allowing for Repeated Measurements
Published on: January 30, 2018
07:11Surgical Techniques for Catheter Placement and 5/6 Nephrectomy in Murine Models of Peritoneal Dialysis
Published on: July 19, 2018
Related Concept Videos
Cavity Drainage and Flashings in Masonry walls
Weep holes, strategically placed at the base of the cavity, are critical for draining accumulated water. These openings are created by leaving head...
Peritoneal Dialysis II: Peritoneal Dialysis Systems and Complications
Peritoneal Dialysis I: Introduction and Procedure
Gravitational Potential Energy for Extended Objects
Pharmacokinetic–Pharmacodynamic Relationship: Duration of Dose-Effect Relationship
Pharmacokinetic–Pharmacodynamic Relationship: Influence of Elimination Half-Life on Effect Duration